Spécifications techniques
|
|
|
Input way
|
Floating and defibrillation protection
|
|
Sampling precision
|
12 bits
|
|
Plomb
|
Standard 12 leads(model:BIP0057)
|
|
Patient leak current
|
<10μA
|
|
Input impedance
|
≥50MΩ
|
|
Frequency response
|
0.05Hz~150Hz(-3dB~+0.4dB)
|
|
Time constant
|
Time constant≥3.2s
|
|
CMRR
|
>60dB, >100dB( Adding filter)
|
|
EMG filter
|
25Hz/35Hz (-3dB)
|
|
AC filter
|
50Hz/60Hz (≥20dB)
|
|
Recording way
|
Thermal printing system
|
|
Specification of recording paper
|
210mm(W)*20m(L) high-speed thermal paper
|
|
Paper speed
|
Auto record: 25mm/s, 50mm/s, error:±5%
Rhythm record: 25mm/s, 50mm/s, error:±5% Manual record:5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error:±5% |
|
Sensitivity selections
|
2.5,5,10,20,40mm/mV, error:±5%. Standard sensitivity is 10mm/mV±0.2mm/mV
|
|
Auto record
|
Record setup according to auto record format and mode, automatically changing leads, measuring and analysis.
|
|
Rhythm record
|
record setup according to rhythm record format and mode, automatically measuring and analysis.
|
|
Manual record
|
record setup according to manual record format, manually changing leads.
|
|
Measurement parameters
|
HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P A*is, QRS A*is, T A*is, R.(V5) , S(V1) , R.(V5)+S(V1).
|
|
Product safety type
|
Class I CF applied part, which includes defibrillation and pacing protection circuit.
|
|
Enduring polarization voltage
|
±500mV
|
|
Noise level
|
≤15μVp-p
|
|
Fuse specification
|
2 pcs φ5*20mm AC time lag; T1.6AL 250V
|
|
Taille
|
340mm(L)*320mm(W)*85mm(H)
|
|
Net Weight
|
3.2Kg
|
• ECG lead wire 1 ensemble
• Limb electrode clamp 1 ensemble
• Chest electrode 1 ensemble
• Power cord 1 morceau
• Ground wire 1 ensemble
• Thermal recording paper 1 ensemble
• User manual 1 ensemble
Reviews
There are no reviews yet.